Interface pressure profile analysis for patellar tendon-bearing socket and hydrostatic socket

Moo, E.K. and Abu Osman, Noor Azuan and Pingguan-Murphy, Belinda and Wan Abas, Wan Abu Bakar and Spence, W.D. and Solomonidis, S.E. (2009) Interface pressure profile analysis for patellar tendon-bearing socket and hydrostatic socket. Acta of Bioengineering and Biomechanics, 11 (4). pp. 37-43. ISSN 1509-409X, DOI 20405814.

[img]
Preview
PDF
Interface_pressure_profile_analysis_for_patellar_tendon-bearing_socket_and_hydrostatic_socket.pdf - Published Version

Download (567kB)
Official URL: http://www.actabio.pwr.wroc.pl/Vol11No4/6.pdf

Abstract

Conventionally, patellar tendon-bearing (PTB) sockets, which need high dexterity of prosthetist, are widely used. Lack of chartered and experienced prosthetist has often led to painful experience of wearing prosthesis and this will in turn deter the patients to wear the prosthesis, which will further aggravate stump shrinkage. Thus, the hydrostatic socket which demands relatively lower level of fabricating skill is proposed to replace the PTB socket in order to produce the equivalent, if not better, quality of support to the amputee patients. Both sockets' pressure profiles are studied and compared using finite element analysis (FEA) software. Three-dimensional models of both sockets were developed using MIMICS software. The analysis results showed that hydrostatic socket did exhibit more uniform pressure profiles than that of PTB socket. PTB socket showed pressure concentration near the proximal brim of the socket and also at the distal fibula. It was also found that the pressure magnitude in hydrostatic socket is relatively lower than that of PTB socket.

Item Type: Article
Funders: UNSPECIFIED
Additional Information: Moo, E K Osman, N A Abu Pingguan-Murphy, B Abas, W A B Wan Spence, W D Solomonidis, S E eng Poland 2009/01/01 00:00 Acta Bioeng Biomech. 2009;11(4):37-43.
Uncontrolled Keywords: Amputation Stumps/physiopathology/surgery, Amputees/rehabilitation, Artificial Limbs, Biomechanics, Finite Element Analysis, Humans, Hydrostatic Pressure, Malaysia, Models, Biological, Patellar Ligament/*physiopathology/surgery, Weight-Bearing/physiology.
Subjects: T Technology > TA Engineering (General). Civil engineering (General)
Divisions: Faculty of Engineering
Depositing User: Mr Jenal S
Date Deposited: 14 Jan 2013 03:50
Last Modified: 19 Oct 2020 02:05
URI: http://eprints.um.edu.my/id/eprint/4427