Wahabi, Janatul M. and Ung, N. M. and Mahdiraji, Ghafour A. and Wong, Jeannie H. D. (2024) Development and characterisation of a plastic scintillator dosemeter in high-energy photon beams. Radiation Protection Dosimetry, 200 (3). pp. 264-273. ISSN 0144-8420, DOI https://doi.org/10.1093/rpd/ncad303.
Full text not available from this repository.Abstract
The radioluminescent (RL) dosemeter is excellent for real-time radiation measurement and can be used in various applications. A plastic scintillator is often the choice sensor because of its size and tissue equivalency. This study aims to characterise a novel plastic scintillator irradiated with high-energy photon beams. An RL dosimetry system was developed using the plastic scintillator. The RL dosimetry system was irradiated using a linear accelerator to characterise the dose linearity, dose rate, energy dependency and depth dose. The developed system showed a linear response toward the dose and dose rate. An energy dependency factor of 1.06 was observed. Depth dose measurement showed a mean deviation of 1.21% from the treatment planning system. The response and characteristics of the plastic scintillator show that it may be used as an alternative in an RL dosimetry system.
Item Type: | Article |
---|---|
Funders: | Department of Clinical Oncology, Faculty of Medicine, Universiti Malaya |
Subjects: | R Medicine > R Medicine (General) |
Divisions: | Faculty of Medicine > Biomedical Imaging Department Universiti Malaya |
Depositing User: | Ms. Juhaida Abd Rahim |
Date Deposited: | 14 Nov 2024 06:46 |
Last Modified: | 14 Nov 2024 06:46 |
URI: | http://eprints.um.edu.my/id/eprint/45944 |